According to the problem analysis result done by TSA, you can directly begin the subsequent troubleshooting process from the corresponding potential root cause.
5-4-1-1 - Use a multimeter to measure the volts from the power inlet & outlet of the 5v transformer.
- Result 1 - Both 24v DC inlet and 5v DC outlet are identified.
Start 5-4-1-2. - Result 2 - 24v DC inlet is identified, but 5v DC outlet is not identified.
Start 5-4-1-3.
- Result 3 - Both 24v DC inlet and 5v DC outlet are not identified.
Start 5-4-1-4.
5-4-1-2 - Reconnect the 5v DC outlet wire and verify the fridge interlock functionality.
- Result 1 -The fridge interlock is restored to regular operation.
Issue closed. (Root cause: 5v DC outlet wire disconnected.) - Result 2 - The fridge interlock is still malfunctioning.
Escalate to TSA for machine swap.
5-4-1-3 - Replace the 5c DC transformer and verify the fridge interlock functionality.
- Result 1 -The fridge interlock is restored to regular operation.
Issue closed. (Root cause: 5v DC transformer malfunction.) - Result 2 - The fridge interlock is still malfunctioning.
Restart 5-4-1-1.
5-4-1-4 - Verify 24v DC outlet from the power supply.
- Result 1 - 24v DC outlet is identified.
Start 5-4-1-5. - Result 2 - 24v DC outlet is not identified.
Start 5-4-1-6.
5-4-1-5 - Reconnect the 24v DC outlet wire (both the power supply side and the 5v transformer side) and verify the fridge interlock functionality.
- Result 1 -The fridge interlock is restored to regular operation.
Issue closed. (Root cause: 24v DC inlet wire disconnected.) - Result 2 - The fridge interlock is still malfunctioning.
Restart 5-4-1-1.
5-4-1-6 - Replace the power supply and verify the fridge interlock functionality.
- Result 1 -The fridge interlock is restored to regular operation.
Issue closed. (Root cause: Power supply malfunction.) - Result 2 - The fridge interlock is still malfunctioning.
Restart 5-4-1-1.